Evaluation of a new re-usable electrode for continuous monitoring of blood PO2 during open-heart surgery.

作者信息

Claremont D J, Pagdin T M

出版信息

J Med Eng Technol. 1985 Jul-Aug;9(4):174-9. doi: 10.3109/03091908509032602.

Abstract

In 46 patients undergoing cardiac surgery blood PO2 was continuously monitored with an in-line oxygen electrode during cardiopulmonary bypass and readings were compared with intermittent measurements of samples on a blood gas analyser. In 35 patients in whom arterial PO2 was monitored, there were rapid changes in PO2 and the mean difference between the in-line electrode and the blood gas analyser was 6.63 kPa (p less than 0.001). Venous PO2 was monitored in 11 patients and, in contrast to the arterial readings, the venous PO2 was relatively stable. The mean difference between the in-line electrode readings and the blood gas analyser was 0.21 kPa (p less than 0.05). The advantages of continuous monitoring of blood PO2 are described and the relative merits of mixed venous PO2 and mixed venous oxygen saturation measurements are discussed.

摘要

在46例接受心脏手术的患者中,在体外循环期间用在线氧电极连续监测血液PO2,并将读数与血气分析仪上样本的间歇性测量结果进行比较。在35例监测动脉PO2的患者中,PO2有快速变化,在线电极与血气分析仪之间的平均差异为6.63 kPa(p<0.001)。在11例患者中监测了静脉PO2,与动脉读数相反,静脉PO2相对稳定。在线电极读数与血气分析仪之间的平均差异为0.21 kPa(p<0.05)。描述了连续监测血液PO2的优点,并讨论了混合静脉PO2和混合静脉血氧饱和度测量的相对优点。