Suppr超能文献

考特尼-米尔斯检测系统中人类肿瘤集落的自动计数

Automated counting of human tumour colonies in the Courtenay-Mills assay system.

作者信息

Verheijen R H, Whelan R D, Feitz W F, Kenemans P, Vooys G P, Herman C J, Hill B T

出版信息

Cell Biol Int Rep. 1985 Oct;9(10):893-9. doi: 10.1016/s0309-1651(85)90110-9.

Abstract

A procedure for using the Omnicon automated image analysis system for counting colonies grown from a human tumour cell line (COLO 205) in the Courtenay-Mills assay is described. This involves the transfer of the agar medium from culture tubes into petri dishes. Comparisons of observer and instrument counts were done on a blinded basis. Run-to-run correlation coefficient was 0.996 for automated counting and the inter-observer correlation coefficient was 0.984. Both assessments showed a linear relationship between the number of cells plated and the number of colonies grown. Automated colony counting is fast, reliable and provides additional information on colony size distribution, not obtainable with manual counting. This automated procedure will greatly facilitate in vitro drug sensitivity evaluation.

摘要

本文描述了一种在考特尼-米尔斯试验中使用Omnicon自动图像分析系统对源自人肿瘤细胞系(COLO 205)的菌落进行计数的方法。这包括将琼脂培养基从培养管转移到培养皿中。观察者计数和仪器计数的比较是在盲态下进行的。自动计数的批间相关系数为0.996,观察者间相关系数为0.984。两种评估均显示接种的细胞数量与生长的菌落数量之间呈线性关系。自动菌落计数快速、可靠,并提供了手动计数无法获得的关于菌落大小分布的额外信息。这种自动化程序将极大地促进体外药物敏感性评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验