Improved estimation of fructosamine, as a measure of glycated serum protein, with the Technicon RA-1000 analyzer.

作者信息

San-Gil F, Schier G M, Moses R G, Gan I E

出版信息

Clin Chem. 1985 Dec;31(12):2005-6.

PMID:4064290
Abstract

We present an improved fructosamine assay for use with the Technicon RA-1000 analyzer. Features of this assay include: specimen throughput of 100 per hour, within-batch CV of 1.5%, between-batch CV of 2.4%, standard curve linear up to 24.9 mmol of fructosamine per liter, and good correlation (r = 0.82) with hemoglobin A1.