Cano-Ortiz Antonio, Sánchez-Ventosa Álvaro, González-Cruces Timoteo, Villalba-González Marta, Aguilar-Salazar Francisco Javier, Prados-Carmona Juan J, Carpena-Torres Carlos, Carracedo Gonzalo, Villarrubia Alberto
Department of Anterior Segment, Cornea and Refractive Surgery, Hospital Arruzafa, 14012 Cordoba, Spain.
Department of Health and Biomedical Sciences, Universidad Loyola Andalucía, 41703 Sevilla, Spain.
J Clin Med. 2025 Jun 23;14(13):4460. doi: 10.3390/jcm14134460.
: This study aimed to evaluate the visual quality and symptomatology of a non-diffractive extended depth-of-focus (EDoF) intraocular lens (IOL), the Elon 877PEY (Medicontur, Zsámbék, Hungary), three months after implantation. : A cross-sectional case series study was conducted, with measurements taken three months post-implantation of the Elon IOL. A total of 56 implanted eyes from 28 patients (mean age: 64.5 ± 9.5 years) were included in the statistical analysis. The variables analyzed to assess the effectiveness of the Elon IOL included high-contrast visual acuity, contrast sensitivity, the defocus curve, and visual symptoms. : Three months after implantation, the mean residual sphere was 0.00 ± 0.33 D, while the mean residual cylinder was -0.25 ± 0.41 D. Without correction, patients achieved monocular decimal visual acuity values of 0.94 ± 0.26 for distance, 0.79 ± 0.17 for intermediate, and 0.58 ± 0.15 for near vision. The mean uncorrected contrast sensitivity was 1.61 ± 0.15 log. The defocus curve showed visual acuity exceeding 0.80 decimal (0.10 logMAR) over a 2.00 D range and above 0.63 decimal (0.20 logMAR) over a 2.50 D range. The most frequently reported symptoms, with mild severity and bothersomeness, were glare, starbursts, halos, and focusing difficulties. : Patients implanted with the Elon IOL achieved satisfactory visual quality at all distances, comparable to outcomes reported for other EDoF IOLs in the scientific literature.
本研究旨在评估非衍射型扩展景深(EDoF)人工晶状体(IOL)Elon 877PEY(匈牙利Zsámbék的Medicontur公司)植入三个月后的视觉质量和症状表现。
开展了一项横断面病例系列研究,在植入Elon人工晶状体三个月后进行测量。对28例患者(平均年龄:64.5±9.5岁)的56只植入眼进行了统计分析。为评估Elon人工晶状体的有效性而分析的变量包括高对比度视力、对比敏感度、散焦曲线和视觉症状。
植入三个月后,平均残余球镜度为0.00±0.33 D,而平均残余柱镜度为-0.25±0.41 D。未经矫正时,患者的单眼小数视力在远距离为0.94±0.26,中距离为0.79±0.17,近距离为0.58±0.15。平均未矫正对比敏感度为1.61±0.15 log。散焦曲线显示,在2.00 D范围内视力超过小数0.80(logMAR 0.10),在2.50 D范围内超过小数0.63(logMAR 0.20)。报告频率最高的症状为眩光、星芒、光晕和聚焦困难,严重程度和困扰程度均较轻。
植入Elon人工晶状体的患者在所有距离均获得了令人满意的视觉质量,与科学文献中报道的其他EDoF人工晶状体的结果相当。