用两种非侵入性方法评估犬的动脉血压。

Evaluation of the arterial blood pressure of dogs by two noninvasive methods.

作者信息

Chalifoux A, Dallaire A, Blais D, Larivière N, Pelletier N

出版信息

Can J Comp Med. 1985 Oct;49(4):419-23.

Abstract

The determination of the arterial blood pressure was done on 12 healthy mixed breed dogs in both the anesthetized and the conscious state, to evaluate two instruments (Doppler flow detector and infrasonde D4000), in their ability to indirectly determine arterial blood pressure. The coefficients of variation were higher with indirect methods when compared with the results obtained by cannulation. These coefficients were lower with the Doppler flow detector. The correlation study showed that both apparatuses were reliable in most situations. The infrasonde D4000 was more accurate than the Doppler in the conscious animals. However the results showed a lack of precision in hypertensive conscious dogs. The diastolic arterial blood pressure was particularly precise in the case of the anesthetized hypotensive dogs. Its sensitivity allowed it to register muscle movement artifacts. The Doppler flow detector showed less variation and was particularly accurate in both anesthetized and conscious hypertensive dogs. Its sensitivity allowed artifact movement sounds to be detected. The Doppler should be used in quiet surroundings or earphones should be worn by the evaluator. Some form of restraint is needed with the use of both instruments. Even if the correlations with the direct arterial blood pressure values were better with the infrasonde D4000, greater variations were found in the individual readings. The Doppler instrument represents in the hands of the investigators a better instrument for routine monitoring of blood pressure in the dog.

摘要

在12只健康的杂种犬处于麻醉状态和清醒状态时,测定其动脉血压,以评估两种仪器(多普勒血流探测器和D4000型次声探头)间接测定动脉血压的能力。与通过插管获得的结果相比,间接方法的变异系数更高。使用多普勒血流探测器时这些系数更低。相关性研究表明,在大多数情况下两种仪器都是可靠的。在清醒动物中,D4000型次声探头比多普勒更准确。然而结果显示清醒的高血压犬缺乏精确性。在麻醉的低血压犬中,舒张压测定尤为精确。其灵敏度使其能够记录肌肉运动伪迹。多普勒血流探测器的变异较小,在麻醉和清醒的高血压犬中都特别准确。其灵敏度使其能够检测到伪迹运动声音。应在安静环境中使用多普勒,或者评估人员应佩戴耳机。使用这两种仪器都需要某种形式的约束。即使与直接动脉血压值的相关性D4000型次声探头更好,但在个体读数中发现了更大的变异。在研究人员手中,多普勒仪器是犬血压常规监测的更好仪器。

相似文献

6
Indirect blood pressure measurement in the dog.
Lab Anim Sci. 1975 Apr;25(2):197-202.
8
Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetised dogs using ACVIM guidelines.
J Small Anim Pract. 2010 Jun;51(6):318-24. doi: 10.1111/j.1748-5827.2010.00938.x. Epub 2010 May 10.

引用本文的文献

1
Fluid Resuscitation for Refractory Hypotension.难治性低血压的液体复苏
Front Vet Sci. 2021 Mar 10;8:621696. doi: 10.3389/fvets.2021.621696. eCollection 2021.

本文引用的文献

3
Indirect mean arterial pressure in the anesthetized dog.
Am J Physiol. 1980 May;238(5):H664-6. doi: 10.1152/ajpheart.1980.238.5.H664.
7
Indirect blood pressure measurement in the dog.
Lab Anim Sci. 1975 Apr;25(2):197-202.
8
Blood pressure measurement in the dog.
J Am Vet Med Assoc. 1977 Aug 15;171(4):364-8.
10
Indirect measurement of instantaneous arterial blood pressure in the rat.大鼠瞬时动脉血压的间接测量
Am J Physiol. 1979 Nov;237(5):H632-7. doi: 10.1152/ajpheart.1979.237.5.H632.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验