人类屈肌腱与人工肌腱相关的力学性能。

The mechanical properties of human flexor tendons in relation to artificial tendons.

作者信息

Pring D J, Amis A A, Coombs R R

出版信息

J Hand Surg Br. 1985 Oct;10(3):331-6. doi: 10.1016/s0266-7681(85)80055-3.

Abstract

The continuation of an unacceptable failure rate with tendon repair or grafting procedures, largely due to adhesions, suggested that an artificial flexor tendon could be an attractive alternative. A literature search found no published data of the mechanical properties of fresh human finger flexor tendons, so a study of the strength and extensibility of 153 tendons was carried out. The bone insertion strength of twenty middle finger tendons was also examined. The results showed that an artificial tendon should have a strength of approximately 1500N, and that it should extend 13% at that load, an elongation of 26mm for a tendon 200mm long. The insertion strength was less than a half of the tendon strength. This data will allow an artificial flexor tendon to be designed with sufficient strength and the correct elastic properties to allow its function to integrate reliably with natural tendons in adjacent fingers.

摘要

肌腱修复或移植手术的失败率一直居高不下,令人难以接受,这主要是由于粘连所致。这表明人工屈肌腱可能是一个有吸引力的替代方案。文献检索未发现有关新鲜人手指屈肌腱力学性能的已发表数据,因此对153根肌腱的强度和伸展性进行了研究。还检查了20根中指肌腱的骨附着强度。结果表明,人工肌腱的强度应约为1500N,在该负荷下应伸长13%,对于一根200mm长的肌腱,伸长为26mm。附着强度不到肌腱强度的一半。这些数据将有助于设计出具有足够强度和正确弹性特性的人工屈肌腱,使其功能能够与相邻手指的天然肌腱可靠地整合。