Dose measurements and calculations of small radiation fields for 9-MV x rays.

作者信息

Arcovito G, Piermattei A, D'Abramo G, Bassi F A

出版信息

Med Phys. 1985 Nov-Dec;12(6):779-84. doi: 10.1118/1.595664.

DOI:10.1118/1.595664
PMID:4079871
Abstract

Measurements of dose distribution for square fields with sizes ranging from 1 X 1 to 30 X 30 cm for a 9-MV x-ray beam from a Neptune 10 linear accelerator, manufactured by CGR, are reported. Special attention was paid to field sizes smaller than 4 X 4 cm, used in radiosurgery techniques. To express the dose-monitor units relationship, total, collimator, and phantom scatter correction factors were obtained by experimental measurements. A strong dependence of these factors on the smallest field sizes (less than 4 X 4 cm) was shown. Measurements of the maximum depth dose dmax, plotted as a function of field size, showed a maximum at about 5 X 5 cm, in good agreement with previous results. dmax was also measured for the smallest fields, demonstrating that the contaminating electron component of the x-ray beam was not responsible for the dmax shift. Analysis of the penumbra width of cross dose distributions, as a function of field sizes, allowed us to postulate that the dmax shift could be due to the phantom scattered photons, which in turn were generated by the collimator scattered photons. Newly derived tissue-maximum ratio and scatter-maximum ratio data were used for dose profile calculations of 2 X 2, 4 X 4, and 10 X 10 cm field sizes. The agreement between experimental and calculated data was found to be +/- 2% within the geometrical edges of the fields and +/- 6% outside of them. A dose profile from the isocenter of a 2 X 2 cm square field moving through a 360 degree rotation arc was obtained and compared with that from the center of a 125I shielded source, as measured by Ling. Advantages and problems relating to the use of x-ray beams from linear accelerators in radiosurgery are discussed.

摘要

相似文献

1
Dose measurements and calculations of small radiation fields for 9-MV x rays.
Med Phys. 1985 Nov-Dec;12(6):779-84. doi: 10.1118/1.595664.
2
Head scatter off-axis for megavoltage x rays.兆伏级X射线的离轴头部散射。
Med Phys. 2003 Apr;30(4):533-43. doi: 10.1118/1.1556609.
3
Buildup region and depth of dose maximum of megavoltage x-ray beams.
Med Phys. 1994 Mar;21(3):411-6. doi: 10.1118/1.597305.
4
The penumbra of a 6-MV x-ray beam as measured by thermoluminescent dosimetry and evaluated using an inverse square root function.通过热释光剂量测定法测量并用反平方根函数评估的6兆伏X射线束的半影。
Med Phys. 1993 Sep-Oct;20(5):1429-38. doi: 10.1118/1.597157.
5
Development and clinical evaluation of an ionization chamber array with 3.5 mm pixel pitch for quality assurance in advanced radiotherapy techniques.用于先进放射治疗技术质量保证的像素间距为3.5毫米的电离室阵列的研制与临床评估
Med Phys. 2016 May;43(5):2283. doi: 10.1118/1.4945414.
6
Experimental measurement of radiological penumbra associated with intermediate energy x-rays (1 MV) and small radiosurgery field sizes.与中能X射线(1兆伏)及小放射外科射野尺寸相关的放射半影的实验测量
Med Phys. 2007 Oct;34(10):3996-4002. doi: 10.1118/1.2775666.
7
Beam characteristics of energy-matched flattening filter free beams.能量匹配无均整器射束的射束特性
Med Phys. 2014 May;41(5):052103. doi: 10.1118/1.4871615.
8
A study of the characteristics of radiation contaminants within a clinically useful photon beam.
Med Phys. 1985 Nov-Dec;12(6):764-8. doi: 10.1118/1.595661.
9
Dosimetry of 24-MV x rays from a linear accelerator.直线加速器产生的24兆伏X射线的剂量测定
Med Phys. 1987 Mar-Apr;14(2):274-81. doi: 10.1118/1.596138.
10
Peripheral dose measurements for 6 and 18 MV photon beams on a linear accelerator with multileaf collimator.使用多叶准直器的直线加速器上6和18兆伏光子束的周边剂量测量。
Med Phys. 2008 Oct;35(10):4396-403. doi: 10.1118/1.2977533.

引用本文的文献

1
Electron FLASH Delivery at Treatment Room Isocenter for Efficient Reversible Conversion of a Clinical LINAC.在治疗室等中心进行电子FLASH 传输,以实现临床 LINAC 的高效可逆转换。
Int J Radiat Oncol Biol Phys. 2021 Jul 1;110(3):872-882. doi: 10.1016/j.ijrobp.2021.01.011. Epub 2021 Jan 11.