Acker H, Pietruschka F, Zierold K
In Vitro Cell Dev Biol. 1985 Jan;21(1):45-8. doi: 10.1007/BF02620913.
Data of the intracellular electrolyte concentration of potassium and chloride in cultured muscle cells measured by x-ray analysis were compared by using the different activity coefficients with intracellular potassium and chloride activities measured with double-barrelled microelectrodes. By using an activity coefficient of 0.6, 95% of the potassium microelectrode measurements are in accordance with the x-ray analysis values, in spite of a scattering of the values. Membrane potential and intracellular potassium values are linearly related. x-ray analysis and ion-sensitive microelectrodes measured the cytoplasmic chloride in the same range. Taking into account known activity coefficients, an error of 25% must be assumed with the intracellular chloride measurements. However, x-ray analysis and ion-sensitive microelectrode investigations are reliable tools to study intracellular potassium and chloride changes, which play an important role in membrane characteristics.