Steinberg Ethan, Wakim Gisele J, El-Hachem Rita, Kumar Manavii
Anesthesia, University of Miami Miller School of Medicine, Jackson Memorial Hospital, Miami, USA.
Anesthesiology and Pain Medicine, American University of Beirut, Beirut, LBN.
Cureus. 2025 Jul 27;17(7):e88850. doi: 10.7759/cureus.88850. eCollection 2025 Jul.
This case report describes the sudden intraoperative ventilation failure in an elderly patient undergoing thyroid lobectomy following intubation with a Medtronic neural integrity monitoring (NIM) endotracheal tube (ETT). Despite confirming appropriate tube positioning and initiating bronchodilator therapy, ventilation could not be restored until the NIM tube was removed and replaced with a standard ETT. This case predates a Class I FDA recall of Medtronic NIM tubes for design-related obstruction, raising the possibility that device malfunction may have contributed to the clinical presentation. This report highlights the importance of maintaining a broad differential diagnosis during airway emergencies and considering tube exchange early in the process.
本病例报告描述了一名老年患者在使用美敦力神经完整性监测(NIM)气管内导管(ETT)插管后行甲状腺叶切除术时术中突然出现通气衰竭的情况。尽管确认了导管位置合适并开始使用支气管扩张剂治疗,但在移除NIM导管并更换为标准ETT之前,通气无法恢复。该病例发生在美敦力NIM导管因设计相关梗阻被美国食品药品监督管理局(FDA)一级召回之前,提示设备故障可能是导致临床表现的原因之一。本报告强调了在气道紧急情况时保持广泛鉴别诊断以及在病程早期考虑更换导管的重要性。