对两种可从手指进行无创、搏动性、校准血压测量的原型设备的评估。

Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a finger.

作者信息

Smith N T, Wesseling K H, de Wit B

出版信息

J Clin Monit. 1985 Jan;1(1):17-29. doi: 10.1007/BF02832685.

Abstract

We evaluated two prototype instruments that measure pulsatile blood pressure continuously and noninvasively and compared the mean arterial pressure obtained from these devices with that obtained invasively in 17 male surgical patients. Each prototype consisted of an infrared photoplethysmograph mounted inside a finger cuff. The cuff was connected to a pressure control valve, which rapidly changed the cuff pressure so as to maintain a null pressure difference across the finger arterial wall. The resultant cuff pressure rapidly tracked the pulsatile intraarterial pressure. The prototypes reproduced absolute pressure, as well as pressure changes, accurately and linearly over a wide range of mean arterial pressures (from 2 to 164 mm Hg), with an average offset error of 0.8 mm Hg (SD +/- 3.8; range, -4.6 to 7.9), a mean scatter error of 5.3 mm Hg (range, 3.6 to 8.6), a mean regression slope of 0.97 (range, 0.79 to 1.22) and a mean correlation coefficient of the regression of 0.96 (range, 0.89 to 0.98). Both prototypes worked satisfactorily on all 17 patients, but not all the time on all patients. In 7 patients, probable arterial spasm prevented measurement of finger blood pressure 12.1% of the time, or 5.4% of the time for all patients. Ninety-six percent of the lost samples occurred with prototype 2, suggesting an instrument-related cause, rather than one related to the principle itself. The prototypes were simple to use and were almost free from artifact. Continuous monitoring for up to 7 hours on a single finger caused no harm to the finger.

摘要

我们评估了两种能够连续、无创测量脉动血压的原型仪器,并将这些设备测得的平均动脉压与17名男性外科患者通过有创方式测得的平均动脉压进行了比较。每个原型仪器都包括一个安装在手指袖带内的红外光电容积脉搏波传感器。该袖带连接到一个压力控制阀,该阀可快速改变袖带压力,以维持手指动脉壁两端的零压力差。由此产生的袖带压力能够快速追踪动脉内的脉动压力。这些原型仪器在很宽的平均动脉压范围内(2至164毫米汞柱)能够准确、线性地再现绝对压力以及压力变化,平均偏移误差为0.8毫米汞柱(标准差±3.8;范围为-4.6至7.9),平均离散误差为5.3毫米汞柱(范围为3.6至8.6),平均回归斜率为0.97(范围为0.79至1.22),回归的平均相关系数为0.96(范围为0.89至0.98)。两种原型仪器在所有17名患者身上都能令人满意地工作,但并非在所有患者身上都能一直正常工作。在7名患者中,可能的动脉痉挛导致手指血压测量失败的时间占12.1%,在所有患者中占5.4%。96%的测量失败样本出现在原型2上,这表明是仪器相关原因,而非原理本身的问题。这些原型仪器使用简单,几乎没有伪迹。在单个手指上连续监测长达7小时对手指没有造成伤害。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验