Atalay Kursat, Kocak Ibrahim, Sayin Nihat, Sanliturk Muhammed Ali, Zirtiloglu Sibel
Kanuni Sultan Süleyman Training and Research Hospital, Eye Clinic, Atakent Mah. Turgut Ozal Blv. Kucukcekmece, 34303, Istanbul, Turkey.
Doc Ophthalmol. 2025 Sep 13. doi: 10.1007/s10633-025-10048-9.
The primary aim of this study was to compare pupillometry measurements obtained from the Sirius™ and AL-Scan™ devices under different lighting conditions. Additionally, the secondary aim was to assess and compare the topographic (keratometric) measurements provided by these systems.
A non-interventional and prospective study was conducted on normal patients aged 18-45 years. Patients who did not smoke and had a spherical equivalent ranging from + 3.00 D to - 5.0 were included. Pupillometry measurements were taken after 1 minute of dark adaptation. Only the right eyes were compared for keratometry. All eyes were compared for pupillometry. We compared the K1 (D), K2 (D), KAvg (D), photopic pupillometry (PP), and mesopic pupillometry (MP) values.
We examined a total of 30 women (66.7%) and 15 men (33.3%). AL Scan showed an average K1 of 42.99 ± 1.50 D, an average K2 of 44.04 ± 1.74, and an average Kavg of 43.58 ± 1.62. Sirius topography yielded the following averages: 42.94 ± 1.51 for K1, 43.90 ± 1.75 for K2, and 43.41 ± 1.61 for Kavg. The average PP and MP measures obtained with the Nidek Al-Scan instrument were 3.97 ± 1.08 mm and 6.04 ± 1.08 mm, respectively. The Sirus device yielded average PP and MP values of 5.05 ± 1.04 mm and 6.01 ± 1.08 mm, respectively. Bland-Altman statistics showed a good degree of correlation between the MP measurements of both devices (p = 0.65). However, K1, K2, KAvg, and PP data showed no significant agreement between both devices (p < 0.05).
In our research, Sirius and Nidek Al-Scan equipment yield comparable MP data; however, PP is inconsistent. Device selection with repeated pupillometry data requires more investigation.
本研究的主要目的是比较在不同光照条件下从Sirius™和AL-Scan™设备获得的瞳孔测量值。此外,次要目的是评估和比较这些系统提供的地形图(角膜曲率)测量值。
对18至45岁的正常患者进行了一项非干预性前瞻性研究。纳入不吸烟且等效球镜度数在+3.00 D至-5.0之间的患者。在暗适应1分钟后进行瞳孔测量。仅比较右眼的角膜曲率。所有眼睛均进行瞳孔测量比较。我们比较了K1(D)、K2(D)、KAvg(D)、明视瞳孔测量(PP)和中视瞳孔测量(MP)值。
我们共检查了30名女性(66.7%)和15名男性(33.3%)。AL Scan显示K1平均为42.99±1.50 D,K2平均为44.04±1.74,KAvg平均为43.58±1.62。Sirius地形图得出的平均值如下:K1为42.94±1.51,K2为43.90±1.75,KAvg为43.41±1.61。使用Nidek Al-Scan仪器获得的平均PP和MP测量值分别为3.97±1.08 mm和6.04±1.08 mm。Sirius设备得出的平均PP和MP值分别为5.05±1.04 mm和6.01±1.08 mm。Bland-Altman统计显示两种设备的MP测量值之间具有良好的相关性(p = 0.65)。然而,K1、K2、KAvg和PP数据显示两种设备之间无显著一致性(p < 0.05)。
在我们的研究中,Sirius和Nidek Al-Scan设备得出的MP数据具有可比性;然而,PP不一致。对于重复瞳孔测量数据的设备选择需要更多研究。