一种新型光学生物测量仪(尼德克AL-Scan)与Sirius及超声生物测量仪测量结果的比较。

Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry.

作者信息

Çağlar Çağatay, Kocamış Sücattin İlker, Demir Emre, Durmuş Mustafa

机构信息

Department of Ophthalmology, Faculty of Medicine, Hitit University, Çorum, Turkey.

Department of Biostatistics, Faculty of Medicine, Hitit University, Çorum, Turkey.

出版信息

Int Ophthalmol. 2017 Jun;37(3):491-498. doi: 10.1007/s10792-016-0284-6. Epub 2016 Jul 8.

Abstract

To investigate the accuracy of the measurements of Nidek AL-Scan by comparing with Sirius (CSO, Florence, Italy), a corneal tomography which also employs the Scheimpflug principle, and a commonly used device, ultrasound biometry (UB) (Aviso A/B, Quantel Medical, MT, USA). Right eyes of 85 healthy volunteers (58 women 27 men) with a mean age of 39.24 ± 14.37 years (range 15-68) were enrolled into this comparative prospective study. Average K 2.4, average K 3.3, CCT (central corneal thickness), WTW (white to white distance), ACD (anterior chamber depth) and AL (axial length) were obtained from the AL-Scan and compared with average SimK, CCT, WTW (horizontal anterior chamber diameter) and ACD obtained from Sirius and also compared with ACD and AL obtained from UB. The statistically significant difference was found between all of the measurements (p < 0.001) except the average keratometry values (K2.4, K3.3, SimK) (p = 0.083). There was a perfect correlation between keratometry, CCT and AL measurements of the devices (ICC = 0.977, 0.954, 0.923, respectively) and there was a strong correlation between the WTW measurements of AL-Scan and Sirius (ICC = 0.865). While ACD parameter of AL-Scan and UB showed a perfect correlation (ICC = 0.977), there was a moderate correlation between AL-Scan and Sirius and also between UB and Sirius (ICC = 0.608 and 0.664, respectively). There was a high correlation between the all measurements, besides ACD, of AL-Scan and Sirius and they can be used interchangeably for average keratometry and WTW confidently. However, ACD and CCT have a broader 95 % LoA (-0.039 to 0.744 and -24.985 to 3.691, respectively). In addition, AL-Scan and UB were in good agreement regarding ACD, while differences in AL measurements of UB and AL-Scan were clinically important (95 % LoA = -0.091 to 0.703). Furthermore, UB and Sirius have a moderate agreement regarding ACD (95 % LoA = -0.047 to 0.680).

摘要

为了通过与Sirius(意大利佛罗伦萨的CSO公司)进行比较来研究Nidek AL-Scan测量的准确性,Sirius是一种同样采用Scheimpflug原理的角膜地形图仪,同时还与一种常用设备超声生物测量仪(UB)(美国蒙大拿州Quantel Medical公司的Aviso A/B型)进行比较。85名健康志愿者(58名女性,27名男性)的右眼被纳入这项比较性前瞻性研究,他们的平均年龄为39.24±14.37岁(年龄范围15 - 68岁)。从AL-Scan获得平均K值(K2.4、K3.3)、中央角膜厚度(CCT)、白到白距离(WTW)、前房深度(ACD)和眼轴长度(AL),并与从Sirius获得的平均SimK值、CCT、WTW(水平前房直径)和ACD进行比较,同时还与从UB获得的ACD和AL进行比较。除了平均角膜曲率值(K2.4、K3.3、SimK)(p = 0.083)外,所有测量值之间均存在统计学显著差异(p < 0.001)。各设备的角膜曲率、CCT和AL测量值之间具有完美相关性(ICC分别为0.977、0.954、0.923),AL-Scan和Sirius的WTW测量值之间具有强相关性(ICC = 0.865)。虽然AL-Scan和UB的ACD参数显示出完美相关性(ICC = 0.977),但AL-Scan与Sirius之间以及UB与Sirius之间存在中度相关性(ICC分别为0.608和0.664)。AL-Scan和Sirius除ACD外的所有测量值之间具有高度相关性,它们可放心地用于平均角膜曲率和WTW的测量,二者可互换使用。然而,ACD和CCT的95%一致性界限较宽(分别为-0.039至0.744和-24.985至3.691)。此外,AL-Scan和UB在ACD方面一致性良好,而UB和AL-Scan的AL测量值差异在临床上具有重要意义(95%一致性界限 = -0.091至0.703)。此外,UB和Sirius在ACD方面具有中度一致性(95%一致性界限 = -0.047至0.680)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验