Hallberg J R, Ibbott G S, Carson P L, Hendee W R
Med Phys. 1977 Nov-Dec;4(6):528-34. doi: 10.1118/1.594344.
The proper evaluation of the accuracy of a new computer program for radiation-therapy dosimetry requires consideration of both the mathematical algorithm used in the program and the performance required in clinical applications. As an example, our evaluation of the irregular-fields dosimetry program currently marketed by SHM for their Rad-8 system is described. The evaluation begins with an explanation of the mathematical computation described. The evaluation begins with an explanation of the mathematical computation method, with emphasis on the points where the calculation differs from previous methods. Next, the procedure for setting up the beam data file is discussed. Finally, a step-by-step procedure is described in which calculated doses are compared with measured doses, using a Varian Clinac-4 with lead flattening filter, and the limits within which a +/- 5% accuracy is attainable are estimated. Some sources of error and areas for possible improvements are mentioned.