McGarrity G J, Coriell L L
Appl Microbiol. 1973 Aug;26(2):167-72. doi: 10.1128/am.26.2.167-172.1973.
A mass airflow cabinet for handling and housing of laboratory rodents has been developed and tested. The unit consists of a high-efficiency particulate air filter and uniform distribution of air at a vertical velocity of 19 cm per s. Animals are maintained without bedding in mesh-bottomed cages that rest on rollers for rotation inside the cabinet. There is an air barrier of 90 cm per s separating the cabinet air from room air. Sampling for airborne bacteria yielded an average of 0.03 colony-forming units (CFU) per ft(3) of air inside the cabinet, whereas 28.8 CFU per ft(3) was simultaneously detected outside the cabinet during housekeeping, a reduction of almost three logs. The efficiency of the air barrier was tested by aerosolization of T3 phage. When phage was aerosolized 5 cm outside the cabinet, no phage could be detected 5 cm inside when the fans were operating; with the fans off an average of 1.6 x 10(4) plaque-forming units (PFU) per ft(3) was detected in six tests. Aerosolization of phage inside the cabinet yielded an average of 9 x 10 PFU per ft(3) outside; an average of 4.1 x 10(6) PFU per ft(3) were detected with the fans not in operation, a reduction of more than four logs. In-use studies on effectiveness showed that the cabinet significantly reduced the incidence of mice originally titer-free to Reo-3 virus. Hemagglutination inhibition antibodies to Reo-3 were detected in 9/22 (42%) mice housed in a conventionally ventilated animal laboratory while no seroconversion was detected in any of 22 mice housed in the mass air flow cabinet in the same laboratory.
一种用于饲养和安置实验用啮齿动物的大容量气流柜已研制成功并经过测试。该装置由一个高效空气过滤器和以每秒19厘米的垂直速度均匀分布的空气组成。动物饲养在底部有网眼的笼子里,不铺垫料,笼子放在滚轮上以便在柜子里转动。有一个每秒90厘米的空气屏障将柜子内的空气与室内空气隔开。对空气中细菌的采样显示,柜子内每立方英尺空气中平均有0.03个菌落形成单位(CFU),而在日常清洁时,柜子外同时检测到每立方英尺28.8个CFU,减少了近三个对数级。通过T3噬菌体雾化来测试空气屏障的效率。当在柜子外5厘米处雾化噬菌体时,风扇运转时在柜子内5厘米处检测不到噬菌体;风扇关闭时,在六次测试中平均每立方英尺检测到1.6×10⁴个噬斑形成单位(PFU)。在柜子内雾化噬菌体时,平均每立方英尺在柜子外检测到9×10个PFU;风扇不运转时平均每立方英尺检测到4.1×10⁶个PFU,减少了四个多对数级。使用效果研究表明,该柜子显著降低了最初对Reo - 3病毒无抗体的小鼠的感染率。在传统通风的动物实验室饲养的22只小鼠中有9只(42%)检测到针对Reo - 3的血凝抑制抗体,而在同一实验室的大容量气流柜中饲养的22只小鼠中未检测到任何血清转化。