Patton J N, Pantridge J F
Br Med J. 1979 Feb 24;1(6162):513-4. doi: 10.1136/bmj.1.6162.513.
The mean current required for ventricular defibrillation was measured and found to be 0.35 +/- SE 0.03 A/kg body weight, which is about one-third of that predicted from animal experiments. There was no apparent correlation between the current required and body weight (r = -0.007 +/- SE 0.213). There is no evidence of need for defibrillators storing more than 400 J.
测量了心室除颤所需的平均电流,发现为0.35±标准误0.03安/千克体重,约为动物实验预测值的三分之一。所需电流与体重之间无明显相关性(r = -0.007±标准误0.213)。没有证据表明需要存储超过400焦能量的除颤器。