Suppr超能文献

一种新型心脏假体:圣犹达医疗心脏瓣膜:体内实验结果

A new cardiac prosthesis: the St. Jude Medical cardiac valve: in vivo results.

作者信息

Emery R W, Mettler E, Nicoloff D M

出版信息

Circulation. 1979 Aug;60(2 Pt 2):48-54. doi: 10.1161/01.cir.60.2.48.

Abstract

The St. Jude Medical cardiac valve prosthesis is a low profile, lightweight, bi-leaflet, central-flow device constructed entirely of pyrolytic carbon. In vivo testing was performed in 24 cows in which the tricuspid valve (n = 14) or mitral valve (n = 10) was replaced by a St. Jude Medical prosthesis with a 26 mm tissue annulus diameter. Eight animals died in the operative period. The 16 surviving animals were sacrificed at 4, 8, 12, or 18 weeks. There were no valve-related deaths nor mechanical valve failures. Anticoagulation was not used in the postoperative period during which time monthly laboratory parameters were obtained including hemoglobin, hematocrit, white cell count, red cell count and indices, lactic acid dehydrogenase, serum haptoglobin, and, additionally at sacrifice, platelet and red cell morphology. Values indicated this prosthesis did not cause hemolysis. At sacrifice the pulmonary arterial tree and peripheral organs showed no evidence of thromboemboli. There was smooth endocardial ingrowth on all valve sewing rings. One valve had thrombus formation on the valve sewing ring. Nine animals were anesthetized prior to sacrifice and underwent open cardiac catheterization. Ventriculography (n = 6) showed minimal transvalvar reflux, and atriograms (n = 6) demonstrated central flow. Transvalvar gradients were 0 to 22 mm Hg (n = 9). One valve of an 18-week tricuspid implant was examined for wear; valve life was calculated to be greater than 2500 years. These data in conjunction with in vitro test results justify clinical trial of the St. Jude Medical prosthesis.

摘要

圣犹达医疗公司的心脏瓣膜假体是一种外形低矮、重量轻、双叶、中心血流的装置,完全由热解碳制成。在24头奶牛身上进行了体内测试,其中14头的三尖瓣或10头的二尖瓣被直径为26毫米组织瓣环的圣犹达医疗假体所替代。8只动物在手术期间死亡。16只存活的动物在4、8、12或18周时被处死。没有与瓣膜相关的死亡,也没有机械瓣膜故障。术后未使用抗凝剂,在此期间每月获取实验室参数,包括血红蛋白、血细胞比容、白细胞计数、红细胞计数及指数、乳酸脱氢酶、血清触珠蛋白,另外在处死时还获取血小板和红细胞形态。数值表明该假体未引起溶血。处死时,肺动脉树和外周器官未显示血栓栓塞迹象。所有瓣膜缝合环上都有平滑的心内膜向内生长。一个瓣膜的缝合环上有血栓形成。9只动物在处死前接受麻醉并进行了心脏导管切开术。心室造影(6例)显示跨瓣反流极小,心房造影(6例)显示中心血流。跨瓣压差为0至22毫米汞柱(9例)。对一个植入18周的三尖瓣瓣膜进行了磨损检查;瓣膜寿命计算得出大于2500年。这些数据与体外测试结果一起证明了圣犹达医疗假体进行临床试验的合理性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验