Blood velocity measurements in human retinal vessels.

作者信息

Tanaka T, Riva C, Ben-Sira B

出版信息

Science. 1974 Nov 29;186(4166):830-1. doi: 10.1126/science.186.4166.830.

Abstract

Laser Doppler velocimetry was used to measure the velocity of blood in human retinal vessels. The mean flow velocities obtained were 1.9 centimeters per second in a retinal vein and 2.2 centimeters per second in a retinal artery. Scattered light from a weak helium-neon laser beam focused on the vessel was detected by a photomultiplier, and the temporal correlation of the intensity fluctuations was measured with a photon counting autocorrelator. Autocorrelation functions for blood flowing through glass capillaries were used for calibration.

摘要

激光多普勒测速仪用于测量人体视网膜血管中的血流速度。所获得的平均流速在视网膜静脉中为每秒1.9厘米,在视网膜动脉中为每秒2.2厘米。聚焦在血管上的弱氦氖激光束的散射光由光电倍增管检测,强度波动的时间相关性用光子计数自相关器测量。流经玻璃毛细管的血液的自相关函数用于校准。