Miller D M
S Afr Med J. 1979 Apr 28;55(18):721-5.
A new co-axial anaesthetic circuit, employing a preferential-flow T-piece which eliminates the need for valves in the circuit, has been designed. Its efficiency has been shown to compare favourably with that of conventional circuits for use in spontaneous respiration and under conditions of controlled ventilation. The proposed system has the advantage of convenience, in that there is no need to disconnect the system from the patient nor from the anaesthetic machine when it is necessary to change from spontaneous to controlled ventilation or should a circle absorber be incorporated within the circuit. The system has been shown to offer lower resistance than the Mapleson A circuit.
一种新型同轴麻醉回路已设计出来,它采用了优先流T形管,无需在回路中使用阀门。已证明其效率与用于自主呼吸和控制通气条件下的传统回路相比具有优势。该系统具有便利性,因为当需要从自主通气转换为控制通气或回路中应包含环路吸收器时,无需将系统与患者或麻醉机断开连接。已证明该系统的阻力比Mapleson A回路低。