Suppr超能文献

羧基血红蛋白的分光光度测量。对康明斯和劳瑟方法的评估。

Spectrophotometric measurement of carboxyhaemoglobin. An evaluation of the method of Commins and Lawther.

作者信息

Lily R E, Cole P V, Hawkins L H

出版信息

Br J Ind Med. 1972 Oct;29(4):454-7. doi: 10.1136/oem.29.4.454.

Abstract

454-457. The spectrophotometric method of Commins and Lawther for the measurement of carboxyhaemoglobin (COHb) concentration was evaluated for both accuracy and sensitivity. The method was calibrated by equilibrating diluted blood samples with known concentrations of carbon monoxide (CO) in air and measuring the resultant COHb saturations. Over the range 0-40% COHb saturation the mean error was 0·67% COHb (SD 0·65). It is considered that the accuracy of the method is more than sufficient for measurement of both the high COHb levels found in clinical situations and the lower COHb saturation resulting from exposure to urban air and cigarette smoking.

摘要

454 - 457。对康明斯和劳瑟用于测量碳氧血红蛋白(COHb)浓度的分光光度法的准确性和灵敏度进行了评估。该方法通过用已知浓度的一氧化碳(CO)平衡稀释的血样并测量所得的COHb饱和度来进行校准。在0 - 40% COHb饱和度范围内,平均误差为0.67% COHb(标准差0.65)。认为该方法的准确性对于测量临床情况下发现的高COHb水平以及因接触城市空气和吸烟导致的较低COHb饱和度而言绰绰有余。