Sanderson J H, Sotheran M F, Stattersfield J P
Br J Ind Med. 1978 Feb;35(1):67-72. doi: 10.1136/oem.35.1.67.
A quick and accurate method of determining the concentration of carboxyhaemoglobin (COHb) in blood has been developed. The method uses a dual wavelength double beam spectrophotometer in the 1st derivative mode, linked to a digital voltmeter (DVM), with the two beams set 3 nm apart around an isobestic point of reduced haemoglobin (Hbred) and carboxyhaemoglobin at 579 nm. The 1st derivative mode measures the slope, and this slope is proportional to the concentration of COHb.
已开发出一种快速准确测定血液中碳氧血红蛋白(COHb)浓度的方法。该方法采用处于一阶导数模式的双波长双光束分光光度计,连接到数字电压表(DVM),两束光在还原血红蛋白(Hbred)和碳氧血红蛋白在579nm处的等吸收点周围设置为相距3nm。一阶导数模式测量斜率,该斜率与COHb的浓度成正比。