Balkcom R M, O'Donnell C M, Amano E
Clin Chem. 1979 Oct;25(10):1831-5.
The dilution/activity curve for the DuPont aca kinetic amylase (EC 3.2.1.1) method is linear to approximately six times the upper limit of the expected range. Total reaction time on the aca is 4.3 min; however, there is a nonlinear rate of change in absorbance for at least the first 8 min, which is consistent with a lag phase. The within-run and day-to-day coefficients of variation were 2.4% at 100 U/L and 3.6% at 114 U/L, respectively. There was no interference by marked hemolysis (hemoglobin 4 g/L) or by glucose concentrations of 10--50 g/L. Comparisons by the split-sample method between the Amylochrome procedure and the aca procedure gave correlation coefficients of 0.992 for serum and 0.989 for urine. A second comparison, between the AutoAnalyzer I method and the aca, gave correlation coefficients of 0.998 for serum and 0.909 for urine. The expected range for serum was 18--106 U/L and for urine 2--33 U/2 h.
杜邦aca动力学淀粉酶(EC 3.2.1.1)方法的稀释/活性曲线在约预期范围上限的六倍内呈线性。aca上的总反应时间为4.3分钟;然而,至少在最初8分钟内吸光度的变化率是非线性的,这与延迟期一致。批内和日间变异系数在100 U/L时分别为2.4%,在114 U/L时为3.6%。显著溶血(血红蛋白4 g/L)或葡萄糖浓度为10 - 50 g/L时无干扰。通过分割样本法对Amylochrome法和aca法进行比较,血清的相关系数为0.992,尿液的相关系数为0.989。第二次比较是自动分析仪I法和aca法之间,血清的相关系数为0.998,尿液的相关系数为0.909。血清的预期范围是18 - 106 U/L,尿液的预期范围是2 - 33 U/2 h。