Physical parameters affecting the visibility of small ventricular septal defects using two-dimensional echocardiography.

作者信息

Jaffe C C, Atkinson P, Taylor K J

出版信息

Invest Radiol. 1979 Mar-Apr;14(2):149-55. doi: 10.1097/00004424-197903000-00007.

Abstract

We report a study showing that two-dimensional echocardiography is capable of visualizing a ventricular septal defect as small as 3 mm. The physical parameters which determine the visibility of an aperture in the myocardium include the presence of anterior refractile tissue, the amount of ultrasound beam focusing and range over which this occurs, the angle of incidence of the ultrasound beam on the septum, and the number of lines of information per degree of sector sweep. These factors are illustrated by in vitro tests. The influence of operator controllable variables in two-dimensional sector scanning echocardiography is discussed.

摘要

我们报告了一项研究,该研究表明二维超声心动图能够可视化小至3毫米的室间隔缺损。决定心肌中孔洞可见性的物理参数包括前方折射组织的存在、超声束聚焦的程度及发生聚焦的范围、超声束与间隔的入射角以及每度扇形扫描的信息线数量。体外试验说明了这些因素。本文还讨论了二维扇形扫描超声心动图中操作者可控变量的影响。