纳氏麻醉气体监测仪的响应时间。

Response time of the Narkotest anesthetic gas monitor.

作者信息

Velazquez J L, Feingold A, Walther P

出版信息

Anesth Analg. 1977 May-Jun;56(3):395-7. doi: 10.1213/00000539-197705000-00016.

Abstract

Response time (RT) of the Narkotest anesthetic gas monitor was measured for 7 inhalation anesthetics, using a 5 L/min fresh gas flow (FGF). Time to 63 percent of maximum response (RT63) and to 95 percent of maximum response (RT95) was directly related to rubber/gas and oil/gas partition coefficients. RT95 ranged from 7 seconds for N2O to 843 seconds for methoxyflurane. RT measured at 0.5 L/min FGF was markedly prolonged over RT at 5 L/min. When the Narkotest was placed on the expiratory limbs of circle-absorber breathing circuits, the difference between the calibrated Narkotest reading (Fn) and the mixed expired circuit concentration (Fc) was shown to depend on the rate of increase of the circuit concentration.

摘要

使用5升/分钟的新鲜气流(FGF),对7种吸入麻醉剂测量了Narkotest麻醉气体监测仪的响应时间(RT)。达到最大响应的63%(RT63)和95%(RT95)的时间与橡胶/气体和油/气体分配系数直接相关。RT95范围从氧化亚氮的7秒到甲氧氟烷的843秒。在0.5升/分钟FGF下测量的RT比在5升/分钟时显著延长。当Narkotest置于循环吸收式呼吸回路的呼气支上时,校准的Narkotest读数(Fn)与混合呼出回路浓度(Fc)之间的差异显示取决于回路浓度的增加速率。