Am J Optom Physiol Opt. 1977 Dec;54(12):834-6. doi: 10.1097/00006324-197712000-00005.
The Goldmann, Mackay-Marg, and Non-contact tonometers were compared on 25 subjects to establish correlations between measurements of intraocular pressure by these instruments. Topical anesthetic was used for the Goldmann and Mackay-Marg readings only. Correlations were high: between the Goldmann and the Non-contact, +0.93; between the Mackay-Marg and the Non-contact, +0.81; and between the Goldmann and the Mackay-Marg, +0.87.
对25名受试者使用戈德曼眼压计、麦凯-马格眼压计和非接触眼压计进行测量,以确定这些仪器所测眼压值之间的相关性。仅在使用戈德曼眼压计和麦凯-马格眼压计测量时使用了表面麻醉剂。相关性很高:戈德曼眼压计与非接触眼压计之间为+0.93;麦凯-马格眼压计与非接触眼压计之间为+0.81;戈德曼眼压计与麦凯-马格眼压计之间为+0.87。