比较治疗青光眼患者的眼压回弹式测量、非接触式眼压测量和 Goldmann 压平眼压测量。

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

机构信息

School of Optometry, Queensland University of Technology, Brisbane, Queensland, Australia.

出版信息

Clin Exp Ophthalmol. 2012 May-Jun;40(4):e163-70. doi: 10.1111/j.1442-9071.2011.02670.x. Epub 2011 Oct 20.

Abstract

BACKGROUND

To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients.

DESIGN

A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic.

PARTICIPANTS OR SAMPLES

One hundred nine (54 males : 55 females) patients including only eyes under medical treatment for glaucoma.

METHODS

Measurement by Goldmann applanation tonometry, iCare rebound tonometry and 7CR non-contact tonometry.

MAIN OUTCOME MEASURES

Intraocular pressure.

RESULTS

There were strong correlations between the intraocular pressure measurements obtained with Goldmann and both the rebound and non-contact tonometers (Spearman r-values ≥ 0.79, P < 0.001). However, there were small, statistically significant differences between the average readings for each tonometer. For the rebound tonometer, the mean intraocular pressure was slightly higher compared with the Goldmann applanation tonometer in the right eyes (P = 0.02), and similar in the left eyes (P = 0.93); however, these differences did not reach statistical significance. The Goldmann correlated measurements from the non-contact tonometer were lower than the average Goldmann reading for both right (P < 0.001) and left (P > 0.01) eyes. The corneal compensated measurements from the non-contact tonometer were significantly higher compared with the other tonometers (P ≤ 0.001).

CONCLUSIONS

The iCare rebound tonometer and the 7CR non-contact tonometer measure intraocular pressure in fundamentally different ways to the Goldmann applanation tonometer. The resulting intraocular pressure values vary between the instruments and will need to be considered when comparing clinical versus home acquired measurements.

摘要

背景

比较 iCare 回弹眼压计和 7CR 非接触眼压计与 Goldmann 压平眼压计在治疗青光眼患者中的眼压读数。

设计

一项在私人三级青光眼诊所进行的前瞻性、横断面研究。

参与者或样本

仅包括接受药物治疗的青光眼患者的 109 只眼(54 名男性:55 名女性)。

方法

Goldmann 压平眼压计、iCare 回弹眼压计和 7CR 非接触眼压计测量。

主要观察指标

眼压。

结果

Goldmann 眼压计与回弹眼压计和非接触眼压计之间的眼压测量值高度相关(Spearman r 值≥0.79,P<0.001)。然而,每种眼压计的平均读数之间存在较小但有统计学意义的差异。对于回弹眼压计,右眼的平均眼压略高于 Goldmann 压平眼压计(P=0.02),左眼相似(P=0.93);然而,这些差异没有达到统计学意义。非接触眼压计的 Goldmann 相关测量值低于右眼(P<0.001)和左眼(P>0.01)的平均 Goldmann 读数。非接触眼压计的角膜补偿测量值明显高于其他眼压计(P≤0.001)。

结论

iCare 回弹眼压计和 7CR 非接触眼压计与 Goldmann 压平眼压计在眼压测量方面的基本原理不同。仪器之间的眼压值不同,在比较临床和家庭获得的测量值时需要考虑这些差异。