MacKay-Marg眼压计、Goldmann眼压计和Perkins眼压计在异常角膜中的比较。

Comparison of MacKay-Marg, Goldmann, and Perkins tonometers in abnormal corneas.

作者信息

McMillan F, Forster R K

出版信息

Arch Ophthalmol. 1975 Jun;93(6):420-4. doi: 10.1001/archopht.1975.01010020434007.

Abstract

Intraocular pressure was determined manometrically in owl monkey eyes with normal and edematous corneas. The pressure was measured with Perkins hand-held, Goldmann, and MacKay-Marg tonometers to compare relative accuracy. At monometric levels of 20, 30, and 40 mm Hg, a statistically significant difference in readings was found between normal and edematous corneas with use of the Perkins and Goldmann applanators, whereas no such difference was found with the MacKay-Marg tonometer. The MacKay-Marg tonometer most closely approximated the true intraocular pressure in edematous corneas.

摘要

用压力测定法测量了角鸮猴正常角膜和水肿角膜眼内的眼压。使用珀金斯手持式眼压计、戈德曼眼压计和麦凯 - 马格眼压计测量眼压,以比较相对准确性。在20、30和40毫米汞柱的压力测定水平下,使用珀金斯眼压计和戈德曼压平眼压计时,正常角膜和水肿角膜之间的读数存在统计学上的显著差异,而使用麦凯 - 马格眼压计时未发现这种差异。麦凯 - 马格眼压计最接近水肿角膜的真实眼内压。