Suppr超能文献

使用LKB 8600反应速率分析仪测定凝血因子(II - VII - X)及高岭土部分凝血活酶时间。

Determination of coagulation factors (II-VII-X) and kaolin partial thromboplastin time by the LKB 8600 Reaction Rate Analyser.

作者信息

Hjorth B, Kjaersgaard E

出版信息

Scand J Clin Lab Invest. 1977 Feb;37(1):15-20.

PMID:616022
Abstract

Testing of a modified Reaction Rate Analyser 8600 in the automatic determination of coagulation time was accomplished with the assays Thrombotest Normotest and Kaolin Partial Thromboplastin Time. Variations in extinctions, from the adding of start reagent until coagulation, were recorded. The coagulation activities, found by the automatic method for the assays Thrombotest and Normotest, were compared with a manual method. Thrombotest showed a good correlation: 0.30 greater than P greater than 0.25 (paired t test). Normotest showed significantly different values by the methods: P less than 0.001 (paired t test). Kaolin Partial Thromboplastin Time (Activated Partial Thromboplastin Time with kaolin as activator) gave significantly lower vales--the difference was 3 sec--compared with a manual method, and the precision was considerably improved.

摘要

采用血栓试验、正常试验和高岭土部分凝血活酶时间测定法,对改良的反应速率分析仪8600自动测定凝血时间进行了测试。记录了从加入起始试剂到凝血过程中的吸光度变化。将血栓试验和正常试验自动测定法得到的凝血活性与手工测定法进行了比较。血栓试验显示出良好的相关性:0.30>P>0.25(配对t检验)。正常试验显示两种方法的结果有显著差异:P<0.001(配对t检验)。高岭土部分凝血活酶时间(以高岭土为激活剂的活化部分凝血活酶时间)与手工测定法相比,结果显著降低——差异为3秒——并且精密度有了很大提高。