Andersen I, Thorsen S
Scand J Clin Lab Invest. 1977 Feb;37(1):7-13.
The application of the Reaction Rate Analyser LKB 8600 (RRA) as an automatic coagulometer is described. The RRA was slightly modified without interfering with its function as an enzyme reaction rate analyser. The endpoint of coagulation was recorded when the increase in absorbance exceeded 0.047 at lambda = 340 nm. The coagulation time was monitored by a counter with automatic print-out or by a recorder. The analytical dispersion, s(x)x, ranged between 0.01 and 0.05. Results (x, y) of the determination of the quantity of plasma coagulation factors (II + VII + X) by RRA and manually by visual recording could be expressed by y = 0.93x + 0.04, r = 0.96, n = 66 (method of Owren & Aas) or by y = 1.00x + 0.02, r = 0.98, n = 49 (Normotest). Similarly, plasma activated partial thromboplastin time (Activated Thrombofax) could be expressed by y = 0.99x + 5.00, r = 0.99, n = 23.
描述了反应速率分析仪LKB 8600(RRA)作为自动凝血仪的应用。对RRA进行了轻微改装,但不影响其作为酶反应速率分析仪的功能。当在λ = 340nm处吸光度增加超过0.047时记录凝血终点。凝血时间由带有自动打印输出的计数器或记录仪进行监测。分析离散度s(x)/x在0.01至0.05之间。通过RRA和手动视觉记录测定血浆凝血因子(II + VII + X)量的结果(x,y)可以表示为y = 0.93x + 0.04,r = 0.96,n = 66(奥伦和阿斯方法)或y = 1.00x + 0.02,r = 0.98,n = 49(正常测试)。同样,血浆活化部分凝血活酶时间(活化血栓形成因子)可以表示为y = 0.99x + 5.00,r = 0.99,n = 23。