Riemer J, Kopperschläger G
Folia Haematol Int Mag Klin Morphol Blutforsch. 1984;111(1):72-83.
An enzyme immunoassay system basing on a competitive method has been developed to measure factor VIII related antigen (F. VIII R:Ag). A sufficient discrimination at low F. VIII R:Ag concentrations was gained. This method appears to be sensitive to 7,8 X 10(-3) U/ml F. VIII R:Ag showing an intraassay coefficient of variation (CV) of 0,11. In comparison to the commonly used Laurell electroimmunodiffusion assay for factor VIII significant less antisera per sample for the enzyme immunoassay technique is necessary.
已开发出一种基于竞争法的酶免疫分析系统来检测因子VIII相关抗原(F. VIII R:Ag)。在低F. VIII R:Ag浓度下获得了足够的区分度。该方法对7.8×10⁻³ U/ml的F. VIII R:Ag似乎很敏感,批内变异系数(CV)为0.11。与常用的检测因子VIII的Laurell电免疫扩散法相比,酶免疫分析技术每个样本所需的抗血清显著减少。