Allison F S
Am J Med Technol. 1983 Sep;49(9):625-32.
As the use of laboratory techniques in diagnostic medicine increase, it became obvious that standardized reference methods and stable controls were essential in order for results obtained in different laboratories to be comparable. Certain procedures and standards have been adopted to provide QC in hematology. Currently procedures include the following: cyanmethemoglobin for HGB determinations; particle counter for RBC and WBC counts; phase microscopy for PLT counts; and packed cell volume for HCT determinations. These serve as acceptable reference methods for multichannel hematology instruments. Instruments that are used must be carefully calibrated. Calibration of the spectrophotometer should be performed using an acceptable standard for hemoglobin. Calibration of the particle counter to determine RBC and WBC counts should be performed using a fresh sample of anticoagulated (EDTA) blood. PLT counting by phase hemacytometry should be used to obtain reference values for automated PLT counters. Centrifuges that are used to obtain microhematocrit values should be calibrated for maximum packing times and times checked against an electric clock. After the instruments have been calibrated, fresh whole blood is measured to determine a target value for each determination. Primary calibration of multichannel instruments is performed by calibrating each parameter to the target value. This is performed initially and then once each week. Two techniques that are commonly used to monitor instrument performance are: 1) the use of at least two levels of commercially prepared control samples to prepare either Levey-Jennings charts or the simple Cusum charts, which detect presymptomatic instrument problems; and/or 2) the use of 500-1,000 patient indices and mean WBC counts analyzed by the XB statistic to establish limits for monitoring instrument performance.
随着诊断医学中实验室技术的应用不断增加,很明显,为了使不同实验室获得的结果具有可比性,标准化的参考方法和稳定的对照至关重要。已经采用了某些程序和标准来提供血液学质量控制。目前的程序包括以下内容:采用氰化高铁血红蛋白法测定血红蛋白;采用血细胞分析仪计数红细胞和白细胞;采用相差显微镜法计数血小板;采用红细胞压积测定法测定红细胞比容。这些可作为多通道血液学仪器可接受的参考方法。所使用的仪器必须仔细校准。应使用可接受的血红蛋白标准品对分光光度计进行校准。应使用新鲜的抗凝(乙二胺四乙酸)血液样本对用于测定红细胞和白细胞计数的血细胞分析仪进行校准。应采用相差血细胞计数法计数血小板以获得自动血小板计数器的参考值。用于获得微量血细胞比容值的离心机应校准最大压实时间,并与电子时钟核对时间。仪器校准后,测量新鲜全血以确定每次测定的目标值。多通道仪器的初次校准是通过将每个参数校准到目标值来进行的。这在最初进行,然后每周进行一次。常用于监测仪器性能的两种技术是:1)使用至少两个水平的商业制备对照样本绘制Levey-Jennings图或简单的累积和图,以检测仪器的症状前问题;和/或2)使用500 - 1000个患者指标和通过XB统计分析的平均白细胞计数来确定监测仪器性能的限值。