Pybus D A, Adams A P
Anaesthesia. 1978 Feb;33(2):178-83. doi: 10.1111/j.1365-2044.1978.tb08346.x.
The control module of the Nuffield (A-P) Anaesthetic Ventilator was found to be an efficient means of automatically controlling the ventilation of the lungs through a bronchoscope during a series of laboratory experiments. The relationships between lung characteristics (compliance and airway resistance) and the driving flow rate through various injectors on the final oxygen concentrations and tidal volumes delivered were determined. A 14 gauge injector provided the largest tidal volumes with the least entrainment of air.
在一系列实验室实验中,发现纳菲尔德(前后位)麻醉呼吸机的控制模块是通过支气管镜自动控制肺部通气的有效手段。确定了肺部特征(顺应性和气道阻力)与通过各种注射器的驱动流速之间的关系,以及最终氧气浓度和输送的潮气量之间的关系。一个14号规格的注射器提供了最大的潮气量,同时夹带的空气最少。