Hinrichs J E, Bandt C L, Smith J A
J Periodontol. 1984 May;55(5):294-8. doi: 10.1902/jop.1984.55.5.294.
This investigation was undertaken to compare the variability in a new Periotron model 6000 to the variability in the previous Periotron model 600 and to the variability of the ninhydrin area method (NAM) for measuring quantities of fluid collected on filter paper strips. A Hamilton microliter syringe was used to deliver normal human serum to filter paper strips. Strips receiving varying quantities were randomly and blindly assigned to Periotron models 6000 and 600 until 10 strips containing each of 0.15, 0.30, 0.45, 0.60, 0.75, 0.90 and 1.05 microliter had been read using model 6000 and 10 strips containing each of 0.15, 0.30 and 0.45 microliter had been read using model 600. All strips were subsequently read using the NAM. The entire procedure was repeated using distilled water containing 0.1% methyl green. Coefficients of variation (CV) were computed for all sets of data. In all cases the CVs for the Periotron 6000 were significantly smaller than those for the Periotron 600 or the NAM. Pooled estimates of CVs (0.15-0.45 microliter) for the Periotron models 6000 and 600 were 0.054 and 0.116, respectively, (P less than or equal to 0.005) with serum and 0.059 and 0.095, respectively, (P less than or equal to 0.05) with distilled water containing 0.1% methyl green. Pooled estimates of CVs (0.15-1.05 microliter) for the Periotron model 6000 and the same strips subsequently measured with the NAM were 0.046 and 0.069, respectively, (P less than or equal to 0.005) with serum and 0.055 and 0.076, respectively, (P less than or equal to 0.001) with distilled water containing 0.1% methyl green.(ABSTRACT TRUNCATED AT 250 WORDS)
本研究旨在比较新型Periotron 6000型与之前的Periotron 600型的变异性,以及茚三酮面积法(NAM)在测量滤纸条上收集的液体量时的变异性。使用汉密尔顿微量注射器将正常人血清注入滤纸条。将接受不同量血清的滤纸条随机且盲法分配给Periotron 6000型和600型,直到使用6000型读取了10条分别含有0.15、0.30、0.45、0.60、0.75、0.90和1.05微升血清的滤纸条,且使用600型读取了10条分别含有0.15、0.30和0.45微升血清的滤纸条。随后使用NAM读取所有滤纸条。使用含有0.1%甲基绿的蒸馏水重复整个过程。计算所有数据集的变异系数(CV)。在所有情况下,Periotron 6000型的CV均显著小于Periotron 600型或NAM的CV。Periotron 6000型和600型(0.15 - 0.45微升)的CV合并估计值,在血清测试中分别为0.054和0.116(P≤0.005),在含有0.1%甲基绿的蒸馏水中分别为0.059和0.095(P≤0.05)。Periotron 6000型(0.15 - 1.05微升)与随后用NAM测量的相同滤纸条的CV合并估计值,在血清测试中分别为0.046和0.069(P≤0.005),在含有0.1%甲基绿的蒸馏水中分别为0.055和0.076(P≤0.001)。(摘要截选至250字)