Bell E F, Rios G R
Crit Care Med. 1983 Aug;11(8):663-7. doi: 10.1097/00003246-198308000-00017.
Two double-walled incubators, the Air Shields C-100 and the Ohio IC, were evaluated for performance characteristics. The Ohio incubator heated more rapidly from room temperature to 36 degrees C, but overshot the preset air temperature and produced greater fluctuation in air temperature due to the operation of the servocontrol system at equilibrium. Neither incubator produced excessive air currents, sound levels, or carbon dioxide accumulation.
对两台双壁培养箱,即空气护盾C - 100和俄亥俄IC培养箱的性能特征进行了评估。俄亥俄培养箱从室温加热到36摄氏度的速度更快,但超过了预设的空气温度,并且由于伺服控制系统在平衡状态下运行,导致空气温度波动更大。两台培养箱均未产生过多的气流、噪音水平或二氧化碳积聚。