Hoffmann R, Wolf R
Rofo. 1984 Oct;141(4):442-7. doi: 10.1055/s-2008-1053166.
Phantom dose measurements were carried out at a mammography unit-type Siemens Mammomat--using LiF thermoluminescence dosimeters. Full calculative corrections were accomplished for the difference between the elementary composition of the phantom material and the human breast. As the results show, the dose required is smaller by about the factor 6 compared to techniques not using intensifying screens. The mean dose lies between 0.07 and 0.31 rd per X-ray picture.
使用LiF热释光剂量计在西门子Mammomat型乳腺摄影设备上进行了体模剂量测量。针对体模材料与人类乳房基本成分之间的差异完成了全面的计算校正。结果表明,与不使用增感屏的技术相比,所需剂量小约6倍。每张X线照片的平均剂量在0.07至0.31拉德之间。