热膜风速仪——一种测定血流速度的方法。II. 与电磁血流计的体内比较。

The hot-film anemometer--a method for blood velocity determination. II. In vivo comparison with the electromagnetic blood flowmeter.

作者信息

Paulsen P K

出版信息

Eur Surg Res. 1980;12(2):149-58. doi: 10.1159/000128120.

Abstract

Using a constant temperature hot-film anemometer and an electromagnetic blood flowmeter, volumetric flows and velocity profiles were registered in the pulmonary artery, ascending aorta, abdominal aorta and superior vena cava of mongrel dogs. The anemometer registered in 3 out of 4 dogs in the ascending aorta and in 4 out of 5 dogs in the pulmonary artery. The flow profile in these two vessels was flat with a slight deviation with the highest velocity nearer to the posterior wall. In the abdominal aorta the flow profile was sinusoid and in the superior vena cava irregular. In 22 simultaneous measurements anemometer mean results were 97 +/- 23% (+/- SD) of flowmeter results and peak results correspondingly 113 +/- 23%. None of these differences were significant. It is stressed that both qualitatively and quantitatively hot-film anemometer results are comparable to electromagnetic flowmeter results. However, certain differences have been demonstrated.

摘要

使用恒温热膜风速仪和电磁血流计,记录杂种犬肺动脉、升主动脉、腹主动脉和上腔静脉的容积流量和速度分布。风速仪在4只犬中的3只升主动脉以及5只犬中的4只肺动脉中记录到数据。这两根血管中的血流分布呈扁平状,有轻微偏差,最高速度更靠近后壁。腹主动脉中的血流分布呈正弦曲线,上腔静脉中的血流分布不规则。在22次同步测量中,风速仪的平均结果为血流计结果的97±23%(±标准差),峰值结果相应为113±23%。这些差异均无统计学意义。需要强调的是,无论是定性还是定量方面,热膜风速仪的结果都与电磁血流计的结果具有可比性。然而,已证明存在某些差异。