Hickman R, Crosier J H, Smith P, Immelman E J, Terblanche J
S Afr Med J. 1975 Feb 22;49(8):261-4.
Recurrent unexplained interference and day-to-day variation were encountered during preliminary studies with a sine-wave electromagnetic flowmeter. Using in vitro techniques, contributory causes were found to be defective leads to the flowmeter cuffs, inadequate contact between cuff and vessel, and failure to use an earth lead. When these faults were corrected, a study was made of the effects of haematocrit, temperature and the comparison of blood vessel or dialysis tubing in vitro. Final in vivo confirmation of the validity of the cuff was obtained by comparison with a cannula probe.
在使用正弦波电磁流量计进行初步研究期间,遇到了反复出现的不明干扰和日常变化。通过体外技术发现,造成这些问题的原因包括流量计袖带的导线有缺陷、袖带与血管之间接触不充分以及未使用接地导线。纠正这些故障后,对血细胞比容、温度的影响以及体外血管或透析管的比较进行了研究。通过与插管探头比较,最终在体内证实了袖带的有效性。