Evans J M, Gray O P, Holland B M, Wardrop C A
Arch Dis Child. 1978 Apr;53(4):330-3. doi: 10.1136/adc.53.4.330.
A simple, rapid, and accurate method for measuring oxygen consumption in infants is described. Expired gas is entrained in a stream of air drawn through a gently applied face mask. Oxygen consumption is derived from (1) gas flow rate, measured with a flowmeter, and (2) the fall in oxygen concentration of the gas mixture, measured with a paramagnetic analyser. A measurement can be made at the bedside within 2 minutes without disturbing the infant. In 17 mature 2-day-old neonates mean oxygen consumption was 6.52 ml/kg per min, SD 0.90, similar to that previously reported using less simple methods,
本文描述了一种用于测量婴儿耗氧量的简单、快速且准确的方法。呼出气体被夹带在通过轻轻佩戴的面罩吸入的气流中。耗氧量可通过以下方式得出:(1)用流量计测量气体流速,以及(2)用顺磁分析仪测量混合气体中氧气浓度的下降。无需干扰婴儿,即可在床边两分钟内完成测量。在17名2日龄成熟新生儿中,平均耗氧量为每分钟6.52毫升/千克,标准差为0.90,与之前使用不太简单的方法所报告的结果相似。