Donaldson N D
J Biomed Eng. 1984 Jul;6(3):237-41. doi: 10.1016/0141-5425(84)90110-9.
An alternative method for electrically joining two rigid terminals within the flexible encapsulant of a prosthesis was sought after welded platinum wires suffered fatigue failure when implanted. The advantage of using spring clips is discussed and fatigue tests to compare spring-clips, made for one of two gold alloys, with welded Platinum or Pt/Ir wires are described. Some physical properties of the superior gold alloy are tabulated.