Acetabular cup groove and pod design and its effect on cement fixation in total hip arthroplasty.

作者信息

Oh I, Sander T W, Treharne R W

出版信息

Clin Orthop Relat Res. 1984 Oct(189):308-12.

PMID:6478703
Abstract

The effect of pods and grooves of a total hip acetabular cup on the cement-cup interfacial fixation strength was studied with five differently designed acetabular cups. The first cup had no pods and no grooves; the second cup had pods and no grooves; the third cup had pods and 1.0-mm-deep grooves; the fourth cup had pods and 1.5-mm-deep grooves; and the fifth cup had pods and 1.5-mm-deep, undercut grooves. All were embedded in cement and in block molds and were torsionally loaded until failure. The cup with no pods and no grooves showed almost no torsional resistance when compared to the cups with pods. While no statistically significant difference was found among the different depths of grooves, the addition of undercutting of grooves tended to give higher resistance to torsional load when compared with the plain grooved cups. Pods and grooves on the cup are prerequisites for adequate fixation; however, deeper grooves not only are unnecessary but pose potential problems in regard to creep, fracture, and wear. Pods on the cup aid in positioning the cup in a concentric manner.

摘要

使用五种不同设计的髋臼杯研究了全髋关节髋臼杯的凸台和凹槽对骨水泥-髋臼杯界面固定强度的影响。第一个髋臼杯没有凸台和凹槽;第二个髋臼杯有凸台但没有凹槽;第三个髋臼杯有凸台和1.0毫米深的凹槽;第四个髋臼杯有凸台和1.5毫米深的凹槽;第五个髋臼杯有凸台和1.5毫米深的底切凹槽。所有髋臼杯均嵌入骨水泥和块状模具中,并进行扭转加载直至失效。与有凸台的髋臼杯相比,没有凸台和凹槽的髋臼杯几乎没有抗扭性。虽然不同深度的凹槽之间没有发现统计学上的显著差异,但与普通凹槽髋臼杯相比,添加凹槽底切往往能提供更高的抗扭载荷能力。髋臼杯上的凸台和凹槽是充分固定的先决条件;然而,更深的凹槽不仅没有必要,而且在蠕变、骨折和磨损方面存在潜在问题。髋臼杯上的凸台有助于将髋臼杯同心定位。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验