Measurement of frequency of ciliary beats of human respiratory epithelium.

作者信息

Yager J, Chen T M, Dulfano M J

出版信息

Chest. 1978 May;73(5):627-33. doi: 10.1378/chest.73.5.627.

Abstract

A method for accurate measurement of beats on ciliated respiratory epithelial cells in vitro is described. The technique allows the frequency of ciliary beats to be recorded directly on minute specimens of human respiratory epithelium obtained by bronchoscopic brushing. A "hanging-drop" preparation of the scraped material is placed on a thin, flat welled slide and is viewed with a phase-contrast microscope at a magnification of x450. Light transmitted through the specimen is deflected in varying intensities due to the beating action of the cilia and is detected by a photomultiplier cell, amplified and transduced as peaks on an oscilloscope. Due to the relatively high magnification, a small group of cilia may be focused upon and their activity measured. Specimens from two groups totaling 53 patients were measured; in the first group the frequency of ciliary beats was measured at 23 degrees C and ranged from 9.1 to 12.9 beats per second with a mean and standard deviation of 11.0 +/- 1.3 beats per second. The second group was measured at 37 degrees C, and values ranged from 10.3 to 16.8 beats per second, with a mean and standard deviation of 13.8 +/- 1.8 beats per second. These values are interpreted to reflect the autonomous frequency of ciliary beats of the isolated respiratory cells. This simple method may be applied to screen for abnormalities of ciliary beating in patients with altered respiratory mucous clearance, as well as to examine the specific effects in vitro of chemicals, drugs, or pollutants on human ciliary activity.

摘要

本文描述了一种在体外精确测量纤毛呼吸上皮细胞搏动的方法。该技术可直接记录通过支气管刷检获得的人类呼吸道上皮微小标本上纤毛搏动的频率。将刮取材料制成“悬滴”标本,置于薄的、有平底孔的载玻片上,用相差显微镜以450倍放大倍数观察。由于纤毛的搏动作用,透过标本的光以不同强度发生偏转,并由光电倍增管检测,放大后在示波器上转换为峰值。由于放大倍数相对较高,可以聚焦一小群纤毛并测量其活动。对两组共53例患者的标本进行了测量;第一组在23摄氏度下测量纤毛搏动频率,范围为每秒9.1至12.9次搏动,平均为每秒11.0±1.3次搏动,标准差为1.3。第二组在37摄氏度下测量,数值范围为每秒10.3至16.8次搏动,平均为每秒13.8±1.8次搏动,标准差为1.8。这些数值被解释为反映分离的呼吸细胞纤毛搏动的自主频率。这种简单的方法可用于筛查呼吸道黏液清除功能改变患者的纤毛搏动异常,也可用于体外研究化学物质、药物或污染物对人类纤毛活动的特定影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验