Suppr超能文献

电极位置对生理性响应心脏起搏器心房感知的影响。

The effect of electrode position on atrial sensing for physiologically responsive cardiac pacemakers.

作者信息

Timmis G C, Westveer D C, Gadowski G, Stewart J R, Gordon S

出版信息

Am Heart J. 1984 Oct;108(4 Pt 1):909-16. doi: 10.1016/0002-8703(84)90454-x.

Abstract

Fully automatic pacing systems rely on accurate identification of spontaneous atrial signals for physiologically responsive pacing. These signals must be discriminated from far-field ventricular activity, which might otherwise be sensed in the atrium. To amplify on the previously reported superiority of bipolar signals and high-impedance circuitry for atrial sensing, we studied the effects of various intraatrial electrode positions on the atrial and ventricular contribution to electrograms recorded in this chamber. Compared with other intraatrial endocardial sites, right atrial signals were greatest in amplitude and slew rate in the appendage (RAA), averaging 3.3 +/- 0.41 mV and 1.15 +/- 0.16 V/sec (mean +/- SEM), respectively. These values were substantially higher than in the low atrium (p less than 0.001 and 0.0005 for amplitude and slew rate, respectively) and the high lateral atrium (p less than 0.05 for slew rate). Appendage atrial electrograms also had significantly higher amplitude and slew rate than far-field R waves recorded here (p less than 0.0001 for both). Additionally, the greatest difference in spectral content between atrial and far-field ventricular signals was also observed in the RAA. Thus, parameters in the domain of both time and frequency identified the RAA as the superior location for atrial sensing. Except for phrenic nerve problems with pacing, the HRA also appears to be a suitable electrode location for sensing. These considerations are germane in light of a growing number of atrial active and passive fixation leads now being employed for physiologic pacing.

摘要

全自动起搏系统依靠准确识别自发心房信号来进行生理反应性起搏。这些信号必须与远场心室活动区分开来,否则可能会在心房被感知到。为了进一步证明先前报道的双极信号和高阻抗电路在心房感知方面的优势,我们研究了不同心房内电极位置对该腔室记录的心电图中心房和心室成分的影响。与其他心房内心内膜部位相比,右心耳(RAA)处的右心房信号幅度和 slew 率最大,平均分别为 3.3±0.41 mV 和 1.15±0.16 V/秒(均值±标准误)。这些值显著高于低位心房(幅度和 slew 率分别 p<0.001 和 0.0005)和高位外侧心房(slew 率 p<0.05)。心耳处的心房心电图幅度和 slew 率也明显高于此处记录的远场 R 波(两者均 p<0.0001)。此外,在 RAA 中也观察到心房和远场心室信号频谱内容的最大差异。因此,时间和频率域的参数都将 RAA 确定为心房感知的优越位置。除了起搏时的膈神经问题外,高位右心房(HRA)似乎也是一个合适的感知电极位置。鉴于现在越来越多的心房主动和被动固定导线被用于生理起搏,这些考虑是相关的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验