Cretti M, Gloor U, Albrecht R, Bürgi H
Schweiz Med Wochenschr. 1984 Oct 27;114(43):1492-4.
The accuracy of blood glucose measurements was evaluated with two different test strip-reflection photometer systems (Reflolux and Reflocheck). Two experienced technicians took 92 measurements in capillary blood at the bedside and the results were compared with a reference value obtained from the same capillary blood sample by a standard automated glucose oxidase method. Both test strips agreed well with the reference value, with high correlation coefficients and with total errors of less than 25%. It remains to be established whether the two methods are as reliable when performed by patients themselves.
使用两种不同的试纸反射光度计系统(Reflolux和Reflocheck)评估血糖测量的准确性。两名经验丰富的技术人员在床边采集了92份毛细血管血样进行测量,并将结果与通过标准自动葡萄糖氧化酶法从同一毛细血管血样获得的参考值进行比较。两种试纸的测量结果与参考值吻合良好,相关系数高,总误差小于25%。这两种方法由患者自行操作时是否同样可靠仍有待确定。