[Determination of low blood glucose levels in newborns with the reflectance photometer (author's transl)].

作者信息

Dominick H C, Humbel R, Bachmann K D, Banauch D, Carstensen C A

出版信息

Geburtshilfe Frauenheilkd. 1979 Aug;39(8):642-8.

PMID:478265
Abstract

The paper reports the examination of a new test strip Reflotest-Hypoglycemie, which was developed specifically for the hypoglycemic range. In 2 series of 110 respectively 120 samples of capillary blood taken from premature and full term infants the blood glucose concentrations were determined using Reflotest-Hypoglycemie and the hexokineasured with the test strip agree well with the results of hexokinase method in a range less than 50 mg/dl. Bilirubin and hematocrit values do not affect the glucose test. Glutathion and ascorbic acid interfere with the test strip method but this error is neglectable because of physiological low concentrations of these substances.

摘要

该论文报告了一种专门为低血糖范围开发的新型试纸Reflotest-Hypoglycemie的检测情况。在分别取自早产儿和足月儿的110份及120份毛细血管血样本的2组实验中,使用Reflotest-Hypoglycemie测定血糖浓度,在低于50mg/dl的范围内,试纸检测的己糖激酶法结果与己糖激酶法的结果吻合良好。胆红素和血细胞比容值不影响葡萄糖检测。谷胱甘肽和抗坏血酸会干扰试纸法,但由于这些物质的生理浓度较低,这种误差可以忽略不计。