Field comparison between two nitrous oxide (N2O) passive monitors and conventional sampling methods.

作者信息

Bishop E C, Hossain M A

出版信息

Am Ind Hyg Assoc J. 1984 Dec;45(12):812-6. doi: 10.1080/15298668491400692.

Abstract

Field tests conducted over the range of 4-330 ppm nitrous oxide in hospital operating rooms as well as dental and veterinary clinics showed a prototype of the Landauer NITROX passive monitor compared more favorably with a Miran 1-A IR gas analyzer than either the Solid State Sensor or bag samples. All methods were tested simultaneously in a chamber designed to produce uniform real-time monitoring.