Dorn U, Kristen H, Resch P
Arch Orthop Trauma Surg (1978). 1978 Apr 18;91(2):101-5. doi: 10.1007/BF00378891.
An image intensifier--70mm camera--fluoroscopy system combined with a conventional X-ray tube is used for routine examination of the infant's hip. The high quality of 85% of these 70mm exposures makes immediate diagnostic interpretation possible. An even higher percentage can be achieved by most exact positioning of the patients. Apart from the radiation protection of the gonades, the X-ray skin dose of the primary beam is reduced by 90-95% compared with full size radiography using film screen combinations. This exposure technique can certainly be used not only for the examination of the infant's hip.
一种与传统X射线管相结合的影像增强器——70毫米相机——荧光透视系统用于婴儿髋关节的常规检查。这些70毫米曝光中85%的高质量使得立即进行诊断性解读成为可能。通过患者最精确的定位可以实现更高的百分比。除了对性腺的辐射防护外,与使用胶片屏组合的全尺寸放射摄影相比,原发射线的X射线皮肤剂量降低了90 - 95%。这种曝光技术当然不仅可用于婴儿髋关节的检查。