Preece J W, Jensen C W
Oral Surg Oral Med Oral Pathol. 1983 Dec;56(6):655-61. doi: 10.1016/0030-4220(83)90085-3.
Speed group "E" dental films were exposed in thirty-five dental x-ray units and processed under rigidly controlled conditions. The exposure, in milliroentgens required to produce an overall film density between 0.85 and 1.05 density units at the 9 mm. step of an aluminum step-wedge, ranged from 94 to 186 mR. The wide range in normalized exposure required to produce a standard density of 1.0 was associated with half-value layer and effective operating kilovoltage in only a general way. The half-value layer of thirty-five dental x-ray units ranged from 1.9 to 2.9 mm. Al, and their effective operating kilovoltages ranged from 62 to 77 kVp when units were set at 70 kVp. The exposure required to produce a specific radiographic density depended largely on the individual characteristics of the x-ray unit used.
E 组感光度牙科胶片在 35 台牙科 X 光机中进行曝光,并在严格控制的条件下冲洗。在铝梯级楔块 9 毫米步长处产生 0.85 至 1.05 密度单位的整体胶片密度所需的曝光量(以毫伦琴为单位)范围为 94 至 186 mR。产生标准密度 1.0 所需的归一化曝光量范围较宽,这仅在一般情况下与半价层和有效工作千伏有关。35 台牙科 X 光机的半价层范围为 1.9 至 2.9 毫米铝,当设备设置为 70 kVp 时,其有效工作千伏范围为 62 至 77 kVp。产生特定放射密度所需的曝光量在很大程度上取决于所使用的 X 光机的个体特性。