Concurrent validity studies of blood pressure instrumentation. The Philadelphia Blood Pressure Project.

作者信息

Barker W F, Hediger M L, Katz S H, Bowers E J

出版信息

Hypertension. 1984 Jan-Feb;6(1):85-91. doi: 10.1161/01.hyp.6.1.85.

Abstract

There is increasing evidence that blood pressure (BP) screening during adolescence may detect early evidence of hypertensive disease. Choice of instrumentation becomes important to maximize the accuracy and interpretive value of several serial readings. We therefore conducted two studies of adolescents, comparing auscultatory (Baumanometer and Random-zero), ultrasonic (Arteriosonde 1216), and infrasonic (Physiometrics SR-2) devices. These instruments were compared serially (Study 1) for older adolescents (n = 48). The Baumanometer, Arteriosonde 1216, and Physiometrics SR-2 were compared both serially and simultaneously (Study 2) for younger and older adolescents (n = 24). In both studies, the order of device presentation was completely counterbalanced. In Study 1, reliability estimates were high (r greater than 0.65), with the exception of the Arteriosonde 1216 diastolic readings, and the Physiometrics SR-2 yielded significantly lower mean diastolic BP readings than the other devices. Diastolic Phase V readings taken with the Random-zero were significantly higher when the Random-zero was presented at the end of the sequence of instrument presentation. This order of presentation effect is probably attributable to the need for overinflation when using the Random-zero device and may cause an overestimation of diastolic BP in the young. In Study 2, again the Arteriosonde 1216 did not perform reliably in recording diastolic BP. Further, the Physiometrics SR-2 device, although reliable, yielded diastolic readings significantly lower (p less than 0.001) than diastolic Phase IV readings taken both serially and simultaneously.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

越来越多的证据表明,青春期进行血压(BP)筛查可能会发现高血压疾病的早期迹象。为了使多次连续测量的准确性和解释价值最大化,仪器的选择变得很重要。因此,我们对青少年进行了两项研究,比较了听诊器(鲍曼血压计和随机零点血压计)、超声波(动脉超声仪1216)和次声(生理测量仪SR - 2)设备。对年龄较大的青少年(n = 48)进行了这些仪器的连续比较(研究1)。对年龄较小和较大的青少年(n = 24)同时进行了鲍曼血压计、动脉超声仪1216和生理测量仪SR - 2的连续和同步比较(研究2)。在两项研究中,设备展示的顺序都进行了完全平衡。在研究1中,除动脉超声仪1216的舒张压读数外,可靠性估计值较高(r大于0.65),并且生理测量仪SR - 2产生的平均舒张压读数明显低于其他设备。当随机零点血压计在仪器展示序列末尾展示时,其测得的舒张期V相读数明显更高。这种展示顺序效应可能归因于使用随机零点设备时需要过度充气,并且可能导致对年轻人舒张压的高估。在研究2中,动脉超声仪1216在记录舒张压方面同样表现不可靠。此外,生理测量仪SR - 2设备虽然可靠,但其产生的舒张压读数明显低于连续和同步测量的舒张期IV相读数(p小于0.001)。(摘要截断于250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验