Suppr超能文献

双叶瓣、倾斜盘瓣和猪主动脉瓣置换物:体内流体动力学特性

Bileaflet, tilting disc and porcine aortic valve substitutes: in vivo hydrodynamic characteristics.

作者信息

Gray R J, Chaux A, Matloff J M, DeRobertis M, Raymond M, Stewart M, Yoganathan A

出版信息

J Am Coll Cardiol. 1984 Feb;3(2 Pt 1):321-7. doi: 10.1016/s0735-1097(84)80015-7.

Abstract

The St. Jude valve is a new bileaflet disc cardiac valve prosthesis designed to avoid some of the hemodynamic drawbacks of other prostheses. The in vivo flow characteristics of the St. Jude aortic valve (42 patients) were studied and compared with those of three other commonly used aortic prostheses. Björk-Shiley (12 patients), Hancock (27 patients) and Carpentier-Edwards (15 patients). The studies, performed 24 to 48 hours after surgery, included measurements at rest and during augmentation of valve flow by infusion of isoproterenol. The mean performance index for valves of all sizes is higher for the St. Jude than for either porcine valve, both at rest and during isoproterenol infusion (p less than 0.05). Utilizing data both at rest and with isoproterenol, the relation of valve flow and mean systolic gradient for each size of St. Jude valve (19 to 25 mm) indicates the occurrence of small increases in gradient (5.3 to 8.2 mm Hg) as valve flow increases, ranging from 161 to 436 ml/systolic X min. A direct comparison of valve flow and gradient data for all size 25 and 23 mm prostheses at rest indicates a tendency for a lower mean systolic gradient in both mechanical valves than in either porcine valve (p = 0.07). With isoproterenol augmentation of valve flow in 25 mm valves, the gradient is less (p less than 0.05), and the effective orifice area and performance index are larger (p less than 0.05) for the St. Jude than for either porcine valve.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

圣犹达瓣膜是一种新型双叶盘式心脏瓣膜假体,旨在避免其他假体的一些血液动力学缺陷。研究了42例植入圣犹达主动脉瓣膜患者的体内血流特性,并与其他三种常用的主动脉假体(12例植入比约克-希利瓣膜、27例植入汉考克瓣膜、15例植入卡朋蒂埃-爱德华兹瓣膜)的血流特性进行比较。这些研究在术后24至48小时进行,包括静息状态下以及通过输注异丙肾上腺素增加瓣膜流量时的测量。在静息状态和输注异丙肾上腺素期间,所有尺寸的圣犹达瓣膜的平均性能指数均高于两种猪瓣膜(p<0.05)。利用静息状态和使用异丙肾上腺素时的数据,每种尺寸(19至25毫米)的圣犹达瓣膜的瓣膜流量与平均收缩期梯度之间的关系表明,随着瓣膜流量从161至436毫升/收缩期×分钟增加,梯度会出现小幅增加(5.3至8.2毫米汞柱)。静息状态下对所有25毫米和23毫米假体的瓣膜流量和梯度数据进行直接比较表明,两种机械瓣膜的平均收缩期梯度均有低于猪瓣膜的趋势(p = 0.07)。在25毫米瓣膜中,使用异丙肾上腺素增加瓣膜流量时,圣犹达瓣膜的梯度更低(p<0.05),有效开口面积和性能指数更大(p<0.05)。(摘要截断于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验