Erbel R, Schweizer P, Krebs W, Langen H J, Meyer J, Effert S
Am J Cardiol. 1984 Feb 1;53(4):590-7. doi: 10.1016/0002-9149(84)90036-5.
The influence of heart rate on left ventricular (LV) volumes and ejection fraction (EF) using 2-dimensional (2-D) echocardiography during atrial pacing was analyzed. The study was performed in 13 normal control subjects, 23 patients with coronary heart disease and 8 patients with dilated cardiomyopathy. An electronic sector scanner (2.25 MHz, 84 degrees) was used. Under constant scanning of the left ventricle, heart rate was increased, in steps of 20 beats/min, from 80 to 140 beats/min. The 2-D echocardiograms were stored on videotape and analyzed off-line. The end-diastolic and end-systolic volumes (EDV and ESV) were determined using a disc method. Stroke volume (SV) and EF were calculated. Constant LV scanning was possible during atrial stimulation, as shown by the analysis of simultaneously recorded 2-D echocardiograms and cineventriculograms at different heart rates, revealing a constant position of the echocardiographic transducer. Simultaneous recordings of cineventriculography and 2-D echocardiography at 80 and 120 beats/min showed that despite differences in absolute values, percent changes of LV volumes and EF determined with both methods were similar. Thus, changes of LV function can be analyzed by 2-D echocardiography. In normal control subjects, an increase in heart rate of 10 beats/min reduced EDV by 4 ml, ESV by 2 ml, SV by 2 ml and EF by 1%, corresponding to percent reductions of 4, 2, 5 and -2%, respectively. In contrast, the absolute decreases in the patients were 6 ml, 1 ml, 5 ml and 2% and the percent changes 2%, 1%, 8% and 5%.(ABSTRACT TRUNCATED AT 250 WORDS)