Edwards F P, Campbell A R
J Clin Pathol. 1984 Apr;37(4):401-8. doi: 10.1136/jcp.37.4.401.
There have been many reports of the relative hazards of formaldehyde, xylene, and hot paraffin wax fumes in pathology laboratories. In contrast, there have been comparatively few describing efficient extraction systems for their removal. When the opportunity presented to design a new laboratory the efficient removal of these fumes was given a high priority. In designing the extraction systems described consideration was given to the physical properties of the fumes and the degree of freedom of use of the bench, with the object of achieving sub-threshold limit values for each area of exposure. The efficiency of removal of the formaldehyde and xylene was measured using infrared spectroscopy.
病理学实验室中有许多关于甲醛、二甲苯和热石蜡油烟雾相对危害的报告。相比之下,描述有效去除这些烟雾的提取系统的报告相对较少。当有机会设计一个新实验室时,高效去除这些烟雾被列为高度优先事项。在设计所描述的提取系统时,考虑了烟雾的物理性质和工作台的使用自由度,目的是使每个暴露区域达到阈限值以下。使用红外光谱法测量甲醛和二甲苯的去除效率。